Error Codes
Error codes listed in this page are for error codes which you find on LCD of logger or PDF report created by logger.
|
错误代码 | LCD显示 | 报告显示 | 描述 | 解决方案 |
004 |
E004 / Err004 | E0004 |
Device reset resulted from power system restart
|
Reconfigure the logger |
006 |
E006 / Err006 | E0006 | Start time error, please configure device again | Reconfigure the logger |
009 |
E009 / Err009 | Start failure due to low battery | Replace battery (multi-use logger) and reconfigure the logger | |
101 |
E101 / Err101 |
Configuration file checksum error, please configure device again |
Reconfigure the logger | |
102 |
E102 / Err102 |
Configuration file error, incorrect device model
|
Reconfigure the logger with correct configuration profile | |
104 |
E104 / Err104 |
Configuration file error, incompatible device version |
||
105 |
E105 / Err105 |
When target device is locked with AccessKey (device password), you need to provide the correct AccessKey when programming the target device, this error occurs when you did not provide a correct AccessKey when programming a locked device. |
Reconfigure the logger with correct AccessKey | |
106 |
E106 / Err106 |
Configuration file, incorrect auto start time |
Reconfigure the logger with correct start time, which must be later than current time |