Tempod 100X

Guide

Configure Device

  • You need to configure Tempod series data loggers before starting a new recording
  • Repeat Start
    • You could choose to enable "Repeat Start" during configuration, then you could start a new recording without reconfiguration, please note configuring device will also calibrate the clock inside logger to ensure the time accuracy, so you are recommended to configure device periodically
    • Repeat start is only applicable to multi-use device

 

Configure device via dragging & dropping configuration file

 

  • Connect logger to USB port, LCD shows "USb" (USB) on successful connection
  • Download configuration file (.cfg format file) from following online utilities
  • Save configuration file to removable storage device "TempSenDisk" create by TempSen USB data logger, no need to open or execute the configuration file
  1. Configure device via TempSen Express

  • Open TempSen Express program
  • Go to "Devices" in left navigation pane
  • Connect loggers to USB port, device LCD shows "USB" on successful connection, and device card will appear in "Devices" page of TempSen Express
  • Select devices you need to configure
  • Select configuration profile you have created
  • Click "Configure Device" button to apply the selected configuration profile to the selected device(s)

 

After configuration

LCD Logger status

Logger still connecting to USB port

Logger removed from USB port

Manual start (with start delay)

Ready for start recording by pressing Start button

Manual start (with temperature condition)

Ready for start recording by pressing Start button

Auto start

Logger will start recording at preset time automatically, no need to press any buttons

Start Recording

  • Press Start button for 3 seconds
  • “REC” on LCD after recording begins

 

Start Start mode Recording begins

Start delay

If logger was programmed with start delay, LCD shows start delay time after starting logger, recording begins after start delay time elapsed.

"REC" shows on LCD

Temperature Condition

If logger was programmed with tempreature conditions, LCD shows ambient temperature after starting logger, recording begins after ambient temperature reaches the preset temperature conditions.

Deploy Devices

  • Except Tempod 90S dry ice data loggers, please DO NOT place the logger body inside environment lower than -30°C to avoid potential damages
  • For loggers with external probes, please ONLY place the probe inside environment lower than -30°C
  • Please be noted that sometimes the surface temperature of the dry ice thermal isolation package could be lower than -30°C

Stop Recording

  • Press Stop button for 3 seconds
  • LCD shows “StoP” (Stop) when recording was successfully stopped

 

 

  • You could not stop a single use data logger within 30 minutes from the time it being started
  • Recording automatically stops as well after memory full

Download Data

Download Device PDF Report

  • Connect logger to USB port of computer, LCD shows "USb" (USB) or USB icon
  • After successful connection, logger will start creating PDF report automatically in seconds
    • LCD shows "PdF" while generating PDF report 
    • LCD shows "PdF" and generation progress in percentage while generating PDF report (ITAG4 series 4.0 or above)
    • USB icon blinks while generating PDF report (ITAG4 SP)
  • PDF report will be generated in removable storage device "'TempSenDisk"

 

 

 

 

Download Data via TempSen Express

  • Open TempSen Express program
  • Go to "Devices" in left navigation pane
  • Connect loggers to USB port, LCD shows "USB" or USB icon after successful connection, and device will appear in "Devices" window as cards
  • Select devices you want to download
  • Click on  on action bar to download data from the selected devices to TempSen Express

 

  • Data file already exists in TempSen Express will not be downloaded again

Error Codes

Please go to Error Codes for more information about error codes you may find on device LCD or PDF report created by loggers

FAQs

General

Can I download data from logger after battery died?

Data in USB data loggers is always retrievable by connecting logger to USB port of computer to download data, even after battery was dead.

 

Do I need to re-program data logger to start new recording after last recording stops?

Multi-use data loggers like Tempod series should be re-configured for new recording after previous recording was stopped, but you could enable Repeat Start during configuration then after current recording stopped, you could start recording immediately by pressing Start button. Please note every time starting a new recording will erase existing data in logger, so always remember to download and save data in advance.

 

What does the lightening icon on LCD mean?

Some  loggers are shipped to customer with power system off status for power saving purpose, once user connects the logger to USB port, the power system will be activated and lightening icon appears on LCD which means logger starts consuming power, it is a remider for to use the logger ASAP.

Battery

How to correctly replace (install) battery on Tempod loggers?
Why does remaining battery drop to low level or empty, or LCD display abnormally after changing battery?

Sometimes after changing or installing new battery, Tempod may not correctly recognize battery capacity, possible errors including battery indicator on LCD shows only 1 bar or even empty, or LCD displays abnormally or blank, this problem results from the power buffer system in data logger, so when it happens, please try to follow steps below to correctly install or replaceable battery and make sure logger could correctly measure battery capacity again.

 

  • Remove battery from logger

  • Press the View/Mark button for several times until LCD changes to BLANK (to exhaust power buffer)

  • Install battery back to the logger

  • Please re-program data logger if Error codes like E009 appears on LCD

AccessKey

What is AccessKey and what does it protect?

AccessKey is a device password designed to protect data logger from unauthorized programming to data logger, so once a data loger is protected with an AccessKey, user needs to provide correct AccessKey when programming this logger.

 

How can I do if I forgot AccessKey?

When you forgot AccessKey of a data logger, please send PDF report downloaded from device and other information we need to verify your information to support@tempsen.com, we will send you the AccessKey of this device by mail after we verified all information we need.

 

 